ProbeSure™ OneStep RT-PCR Master Mix is a 2x master mix for one-step reverse transcription and PCR from RNA template with hydrolysis probe-based assays, in endpoint or real-time mode. This user guide covers ROX options, primer and probe concentrations, RNA handling, a 10 µL reaction setup, the thermal cycling protocol with an RT step, signal detection and data analysis.
ProbeSure OneStep RT-PCR at a glance
- Applications: one-step RT-PCR from RNA with hydrolysis probes, endpoint or real-time
- Compatible probes: TaqMan™, ZEN™ and BHQ™/BHQplus™ probes
- ROX options: no ROX, low (25 nM) or high (500 nM)
- Pack sizes: 1 mL to 250 mL, 2x concentration
- Typical concentrations: primers 300-900 nM, probes 100-200 nM
- Thermal cycling: RT at 50 °C for 10-30 min, 95 °C for 10-15 min, then 35-45 cycles of 95 °C and 57-65 °C
- Storage: -20 °C / -80 °C for two years, or 4 °C for two weeks protected from light
Working with DNA instead of RNA? See the ProbeSure™ Master Mix user guide.
1. PRODUCT DETAILS

2. DESCRIPTION
ProbeSure OneStep RT-PCR Master Mix is designed for highly specific and sensitive one-step reverse transcriptase (RT) fluorescently-reporting PCR using RNA template. The master mix includes optimised components which allow reverse transcription and subsequent PCR amplification to take place in the same reaction well. ProbeSure OneStep RT-PCR Master Mix may be used in endpoint or real-time modes.
ProbeSure OneStep RT-PCR Master Mix is designed for use in PCR genotyping applications with hydrolysis probe-based assays. It can be used with a variety of fluorogenic probe types including TaqMan™, ZEN™ probes and BHQ™/BHQplus™ probes. It is not intended for use as a general PCR master mix or for HRM applications.
ProbeSure OneStep RT-PCR Master Mix contains all the necessary components to carry out the reaction (except RNA and primers/probes) and is supplied at 2x concentration. It is formulated with dTTP as this improves reaction sensitivity and efficiency when compared to mixes containing dUTP.
3. STORAGE AND SHELF LIFE
ProbeSure OneStep RT-PCR Master Mix is shipped on blue ice. Upon arrival, store at -20˚C/-80˚C (stable for two years); multiple freeze / thaw cycles are not recommended. The mix can also be stored at 4˚C for two weeks (protected from light).
4. SAFETY WARNINGS AND PRECAUTIONS
This product should only be handled by trained laboratory personnel. It is advisable to wear suitable personal protective equipment (PPE) when using the product. In case of contact with skin or eyes, wash immediately with water.
5. KIT COMPONENTS
ProbeSure OneStep RT-PCR Master Mix (supplied at 2x concentration), containing a specifically engineered Taq polymerase, dNTPs, buffer, performance enhancers, MgCl2, passive reference dye (ROX) and Reverse Transcriptase (RT) (200x).
REQUIRED COMPONENTS
- Fluorescent plate reader or qPCR machine capable of reading the fluorophores used
- PCR plate or equivalent and appropriate optically clear seal
- Template RNA
- PCR-grade water
- Primers and fluorogenic hydrolysis probes (or both combined into an assay)
6. ROX COMPATIBILITY
ProbeSure OneStep RT-PCR Master Mix is supplied without ROX, or with low or high ROX levels. Please ensure compatibility between the ROX level of the master mix and the qPCR instrument. Should you require further assistance, please contact your qPCR instrument/plate reader manufacturer, or contact 3CR Bioscience’s Technical Support team.
If a fluorescent plate reader is used instead of a qPCR machine, it is recommended that the high ROX version of the ProbeSure OneStep RT-PCR Master Mix is used.
7. PRIMERS AND PROBES
The optimal primer and fluorogenic hydrolysis probe concentrations should be determined empirically. Primer concentrations of 300 – 900 nM and probe concentrations of 100 – 200 nM are generally suitable for most applications.
8. RNA QUALITY AND QUANTITY
RNA is highly susceptible to ubiquitous RNases. Care should be taken when handling the samples. An empirical optimisation of RNA concentration should be carried out by testing a sample dilution range.
9. CONTROLS
To improve confidence in the data, control samples should be used on the PCR plate in addition to the test samples. Negative controls (no-template controls, or NTCs) should always be used and consist of the same buffer used to hydrate the RNA samples, dispensed into several wells of the PCR plate. Positive controls can also be used, if available, and should consist of RNA samples of known sequence.
When viewing the data, NTCs should show no amplification, giving confidence that any amplification observed is real. Any amplification observed in the NTC wells would indicate contamination or non-specific amplification. The positive control samples should amplify as expected for their sequence.
10. REACTION PROCEDURE
A. ARRAYING TEMPLATE RNA
A liquid-handling system appropriate to the number of RNA samples to array should be used. RNA samples can also be arrayed manually if working with a low number of samples.
B. REACTION ASSEMBLY
Ensure the components are defrosted thoroughly, mix and centrifuge briefly prior to use. The enzymes used in ProbeSure OneStep RT-PCR Master Mix are modified such that they are completely inactive at ambient temperature. Such inactivation allows bench top reaction assembly without leading to primer-dimer issues and other side reactions. The enzymes are reactivated during the initial stage of thermal cycling.
| COMPONENT | FINAL CONCENTRATION | VOLUME FOR 10 µL REACTION (µL) |
| ProbeSure OneStep RT-PCR Master Mix (2x) | 1x | 5 |
| Primer(s) | 300-900 nM | Variable |
| Probe(s) | 100-200 nM | Variable |
| RNA template | Variable | Variable |
| Nuclease-free water | – | To 10 µL |
| TOTAL | – | 10 µL |
A total reaction mix should always be made to eliminate well-to-well variation of component concentrations. Assemble the total reaction mix using Table 1 as a guide.
ProbeSure OneStep RT-PCR Master Mix can be used in reaction volumes of any size, including volumes (below 1 µL) in the appropriate PCR plate. It is important that ProbeSure OneStep RT-PCR Master Mix is used at a final concentration of 1x.
C. TOTAL REACTION MIX DISPENSING & PLATE SEALING
The total reaction mix must now be dispensed into the PCR plate wells. As with DNA dispensing, use a liquid-handling system that is appropriate to the scale of the work. Once the total reaction mix has been dispensed, the PCR plate must be sealed with an optically clear seal and centrifuged to ensure all components are at the bottom of the wells.
ProbeSure OneStep RT-PCR Master Mix can be used with any standard Peltier-based or water-bath thermal cycler. If used in real-time mode, any appropriate qPCR instrument may be used, and the fluorescent data captured at each cycle. Use the thermal cycling protocol detailed in Table 2.
D. THERMAL CYCLING
Place the plate on a thermal cycler or qPCR instrument and carry out the thermal cycling step. The thermal cycling protocol used will vary with probe type but a guide to appropriate conditions is shown in Table 2.


E. FLUORESCENT SIGNAL DETECTION
After thermal cycling is complete, the fluorescent signal is detected and assessed by reading the plate with a fluorescent plate reader or a qPCR instrument in endpoint or real-time modes. If sufficiently defined genotype clusters are not obtained after the initial thermal cycling protocol, the plate should be cycled for an additional three cycles using the conditions detailed in Table 3 and read/analysed again. The additional cycling can be repeated until tight and well separated clusters are observed, though this is rarely required.
If analysing in real-time, data acquisition should be carried out during the annealing/extension step of each cycle.
F. INTERPRETATION OF DATA
Endpoint data analysis and interpretation can be done using cluster analysis software, a qPCR instrument or alternatively can be carried out in Microsoft Excel.
ROX passive reference dye can also be used to eliminate the effect of well-to-well liquid volume differences from the resultant cluster plot data. The inclusion of a passive reference leads to tighter clustering and, as a result, more accurate scoring of data.
Real-time data analysis is carried out using the qPCR instrument software.
11. ORDERING INFORMATION
For ordering details, please visit www.3crbio.com.
12. SUPPORT
If you require any support with the use of ProbeSure OneStep RT-PCR Master Mix or other 3CR Bioscience products, please contact our Technical Support team on support@3crbio.com.
13. LEGAL INFORMATION
For Research Use Only. Not for use in diagnostic procedures.
3CR Bioscience disclaims all warranties with respect to this documentation.
The purchase of this product conveys to the purchaser the limited, non-transferable right to use the purchased amount of the product only to perform internal research for the sole benefit of the purchaser. No right to resell this product or any of its components is conveyed expressly or by implication. This product is for internal research purposes only and is not for use in commercial applications of any kind, including, without limitation, quality control and commercial services such as reporting the results of purchaser’s activities for a fee or other form of consideration.
TaqMan™ is a trademark of Roche Molecular Systems, Inc.
Zen™ Probes is a trademark of Integrated DNA Technologies, Inc.
BHQ™ Probes and BHQplus™ Probes are trademarks of LGC Biosearch Technologies
©2023 3CR Bioscience Ltd. All rights reserved. Intended for molecular biology applications. This product is not intended for the diagnosis, prevention, or treatment of a disease.