ProbeSure™ Master Mix is a 2x master mix for hydrolysis probe-based PCR assays with endpoint or real-time detection, used for SNP and Indel genotyping and for detecting sequences of interest. This user guide covers ROX options, primer and probe concentrations, DNA input, reaction setup, thermal cycling, fluorescence reading and how to interpret endpoint and real-time data.
ProbeSure at a glance
- Applications: SNP and Indel genotyping and sequence detection with hydrolysis probes, endpoint or real-time
- Compatible probes: TaqMan®, ZEN™ and BHQ™/BHQplus™ probes
- ROX options: no ROX, low (25 nM) or high (500 nM)
- Pack sizes: 2.5 mL to 1,000 mL (500 to 200,000 reactions at 10 µL)
- Typical concentrations: primers 300-900 nM, probes 100-200 nM, 1-10 ng gDNA per well
- Thermal cycling: 95 °C for 10-15 min, then 35-45 cycles of 95 °C and 57-65 °C
- Storage: -20 °C / -80 °C for two years, or 4 °C for two weeks protected from light
1. PRODUCT DETAILS

2. DESCRIPTION
ProbeSure Master Mix is designed for use in PCR applications with hydrolysis probe-based assays with either endpoint or real-time detection. It is formulated for detection of Single Nucleotide Polymorphisms (SNPs) and insertion/deletions (Indels), or detection of sequences of interest.
ProbeSure Master Mix can be used with a variety of fluorogenic probe chemistries including TaqMan®, ZEN™ probes and BHQ™/BHQplus™ probes. It is not intended for use as a master mix for general PCR applications nor for HRM applications.
ProbeSure Master Mix contains all the necessary components to carry out the reaction (except template DNA and primers/probes) and is supplied at 2x concentration. It is formulated with dTTP as this improves reaction sensitivity and efficiency when compared to mixes containing dUTP.
3. STORAGE AND SHELF LIFE
ProbeSure Master Mix is shipped on blue ice. Upon arrival, store at -20˚C/-80˚C (stable for two years); multiple freeze/thaw cycles are not recommended. ProbeSure Master Mix can be aliquoted into light-protective tubes to reduce the need for repeated freeze-thaw cycles. The mix can also be stored at 4˚C for two weeks (protected from light).
4. SAFETY WARNINGS AND PRECAUTIONS
This product should only be handled by trained laboratory personnel. It is advisable to wear suitable personal protective equipment (PPE) when using the product. In case of contact with skin or eyes, wash immediately with water.
5. KIT COMPONENTS
ProbeSure Master Mix (supplied at 2x concentration), containing a specifically engineered Taq polymerase, dNTPs, buffer, performance enhancers, MgCl2 and the passive reference dye (ROX).
REQUIRED COMPONENTS
- Fluorescent plate reader or qPCR instrument capable of reading the fluorophores used in the intended assay.
- PCR plate or equivalent and appropriate optically clear seal
- PCR-grade water
- Primers and probes (or both combined into an assay)
- Template DNA
6. ROX COMPATIBILITY
ProbeSure Master Mix is supplied without ROX, or with low or high ROX levels. Please ensure compatibility between the ROX level of the master mix and the qPCR instrument; should you require further assistance, please contact the manufacturer of your qPCR instrument or plate reader or contact 3CR Bioscience’s Technical Support team.
If a fluorescent plate reader is used instead of a qPCR instrument, it is recommended that the high ROX version of the ProbeSure Master Mix is used.
7. PRIMERS AND PROBES
The optimal primer and probe concentrations should be determined empirically. Primer concentrations of 300 – 900 nM and probe concentrations of 100 – 200 nM are generally suitable for most applications.
8. DNA QUALITY AND QUANTITY
It is recommended to use 1-10 ng of gDNA per reaction well, though this will vary with organism genome size (large genomes will require a proportionately larger DNA mass). For optimal results, purified and well-normalised DNA samples should be used. However, when using ProbeSure Master Mix in high throughput, purified DNA is often not commercially practical. ProbeSure Master Mix contains inhibitor resistant components and so will generally work well with DNA that has been crudely extracted, but such samples should be tested before commencing large scale work. Empirical optimisation of DNA concentration by testing a sample dilution range test is the most sensible approach.
9. CONTROLS
To improve confidence in the genotyping data, control samples should be used on the PCR plate in addition to the test samples. Negative controls (no-template controls, or NTCs) should always be used and consist of the same buffer used to hydrate the DNA samples dispensed into several wells of the PCR plate. Positive controls can also be used, if available, and should consist of DNA samples of known genotypes.
When viewing the genotyping data, NTCs should show no amplification and remain around the origin of the cluster plot (see Figure 1a), giving confidence that any amplification observed is real. Any amplification observed in the NTC wells would indicate contamination or non-specific amplification. The positive control samples should cluster in the expected regions for their genotype.
10. REACTION PROCEDURE
A. ARRAYING TEMPLATE DNA
A liquid-handling system appropriate to the number of DNA samples to array should be used. DNA samples can also be arrayed manually if working with a low number of samples.
ProbeSure Master Mix can be used with hydrated or dry DNA samples. Both approaches work equally well but have practical advantages and disadvantages. If low numbers of samples are to be genotyped, it is not worth drying the DNA samples.
However, if high numbers of samples are to be genotyped in one run, drying the DNA samples can improve the resulting data. Hydrated DNA arrayed in a PCR plate will quickly begin to evaporate differentially across the plate (samples near the edges evaporate more quickly than those in the middle). Variation in DNA volumes across the plate will lead to variation in the final reaction concentrations, causing sub-optimal results. For this reason, when drying the DNA into the plate wells the user must ensure that the DNA has been dried to completion. Dried DNA samples will be stable long term at ambient temperature.
If a very small reaction volume is to be used (for example 1.0 µL total volume), it might not be possible to accurately dispense 0.5 µL of DNA and 0.5 µL of total reaction mix. In this example, drying the sample would allow a more realistic 1.0 µL of total reaction mix to be dispensed to the well.
To dry the DNA, once dispensed into a PCR plate, the plate should be centrifuged to ensure the samples are in the bottom of the wells and placed in a laboratory fan oven for one hour at around 55˚C, or until the samples have visibly dried. When assembling the total reaction mix, water must be added in the correct proportion to account for the missing volume of the DNA template.
B. REACTION ASSEMBLY
Ensure the components are defrosted thoroughly, mix and centrifuge briefly prior to use. The enzymes used in ProbeSure Master Mix are modified such that they are completely inactive at ambient temperature. Such inactivation allows bench top reaction assembly without leading to primer-dimer issues and other side reactions. The enzymes are reactivated during the initial stage of thermal cycling.
| COMPONENT | FINAL CONCENTRATION | VOLUME FOR 10 µL REACTION (µL) |
| ProbeSure MM (2x) | 1x | 5 |
| Forward primer(s) | 300-900 nM | Variable |
| Reverse primer | 300-900 nM | Variable |
| Probe 1 | 100-200 nM | Variable |
| Probe 2 | 100-200 nM | Variable |
| DNA template1 | Variable | Variable |
| Nuclease-free water | – | To 10 µL |
| TOTAL | – | 10 µL |
1Final concentration of cDNA 0.1 pg/µL – 10 ng/µL; gDNA 10 pg/µL – 10 ng/µL
A total reaction mix should always be made to eliminate well-to-well variation of component concentrations. Assemble the total reaction mix using Table 1 as a guide.
ProbeSure Master Mix can be used in reaction volumes of any size, including volumes below 1.0 µL in the appropriate PCR plate. It is important that ProbeSure Master Mix is used at a final concentration of 1x.
C. TOTAL REACTION MIX DISPENSING & PLATE SEALING
The total reaction mix must now be dispensed into the PCR plate wells. As with DNA dispensing, use a liquid handling system that is appropriate to the scale of the work. Once the total reaction mix has been dispensed, the PCR plate must be sealed with an optically clear seal and centrifuged to ensure all components are at the bottom of the wells.
D. THERMAL CYCLING
Place the plate on a thermal cycler or qPCR instrument and carry out the thermal cycling step. The thermal cycling protocol used will vary with probe type but a guide to appropriate conditions is shown in Table 2.

E. FLUORESCENT SIGNAL DETECTION
After thermal cycling is complete, the fluorescent signal data should be collected using an appropriate fluorescent plate reader or qPCR machine in endpoint mode.
If the genotype clusters are not sufficiently defined after running the initial thermal cycling protocol, the plate should be cycled for an additional three cycles (see Table 3) then the fluorescent signal data collected again. The additional cycling/data analysis can be repeated until tight and well-separated clusters are observed.

F. INTERPRETATION OF DATA
Endpoint data analysis and interpretation can be carried out using cluster analysis software or alternatively can be carried out in Microsoft Excel (see Figure 1a).
ROX passive reference dye can also be used to eliminate the effect of well-to-well liquid volume differences from the resulting cluster plot data. The inclusion of a passive reference leads to tighter clustering and, as a result, more accurate scoring of data.
Depending on the application, data can be collected in real-time and viewed as in the example shown in Figure 1b.
11. ORDERING INFORMATION
For ordering details, please visit www.3crbio.com.
12. SUPPORT
If you require any support with the use of ProbeSure Master Mix or other 3CR Bioscience products, please contact our Technical Support team on support@3crbio.com.
13. LEGAL INFORMATION
For Research Use Only. Not for use in diagnostic procedures.
3CR Bioscience disclaims all warranties with respect to this documentation.
The purchase of this product conveys to the purchaser the limited, non-transferable right to use the purchased amount of the product only to perform internal research for the sole benefit of the purchaser. No right to resell this product or any of its components is conveyed expressly or by implication. This product is for internal research purposes only and is not for use in commercial applications of any kind, including, without limitation, quality control and commercial services such as reporting the results of purchaser’s activities for a fee or other form of consideration.
TaqMan™ is a trademark of Thermo Fisher Scientific Inc
Zen™ Probes is a trademark of Integrated DNA Technologies, Inc.
BHQ™ probes and BHQ™ Plus are trademarks of LGC Biosearch Technologies
©2023 3CR Bioscience Ltd. All rights reserved. Intended for molecular biology applications. This product is not intended for the diagnosis, prevention, or treatment of a disease.